Dimension Total Prostate Specific Antigen (TPSA) Flex Reagent Cartridge (RF451). Siemens, Newark, DE 19714

FDA device recall Z-1606-2008 · posted 2008-06-28 · Terminated

Recall details

Recalling firm
Siemens Healthcare Diagnostics Inc.
Reason for recall
False Results: The product may exhibit falsely elevated results. The affected lots demonstrate sporadic variability in recovery of calibrator and quality control; therefore this issue may not be readily detected upon calibration or during routine processing of Quality Control.
Action taken
The recalling firm issued a letter (Urgent Field Safety Notice) dated 3/25/08 to inform their customers to discontinue use of any remaining inventory, retest using an alternate lot for those samples from post-prostatectomy patients which recovered positive TPSA results less than 1.0 ng/nL. The firm has requested that you call the them for no-charge replacement product at 1-800-441-9250 and/or for technical assistance.
Root cause
Device Design
Status
Terminated
Product code
MTF
Quantity affected
1631 cartons
Distribution
The product was shipped to medical facilities nationwide. The product was also shipped to Germany, Singapore, Canada, Japan and AE
Date initiated
2008-03-24
Date posted
2008-06-28
Date terminated
2008-10-15
Location
Newark, DE