Dimension(R) IRON Calibrator (DC85), Siemens material number 10445010. The Dimension(R) IRON Calibrator is an in vitro d

FDA device recall Z-0745-2013 · posted 2013-01-29 · Terminated

Recall details

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Reason for recall
Siemens Healthcare Diagnostics has received complaints of IRON calibrations outside of acceptable calibration slope (m) values (0.97 - 1.03) when using IRON calibrator lots 2DN060 and 2FN063. Lots 2GN068 and 2JN074 may also be impacted but no complaints have been received.
Action taken
Siemens sent a "CUSTOMER NOTIFICATION" letter dated December 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact Siemens Technical Solutions Center at 1-800-441-9250 or your local Siemens Healthcare Diagnostics representative for assistance with this issue.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
JIS
Quantity affected
5999
Distribution
Nationwide Distribution.
Date initiated
2012-12-20
Date posted
2013-01-29
Date terminated
2014-01-08
Location
Newark, DE