Dimension Integrated Chemistry Systems intended to measure a variety of analytes in human body fluids.
FDA device recall Z-2133-2014 · posted 2014-07-30 · Terminated
Recall details
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Reason for recall
- There is a misassembly issue with the photometer installed on Dimension Integrated Chemistry Systems instruments. Through internal investigation by Siemens Healthcare Diagnostics has confirmed that a limited number of Dimension photometers were misassembled. These photometers are performing within specifications, but may lead to a shortened life of the source lamp.
- Action taken
- An urgent field safety notice, dated May 2014, was sent to users to inform them of the issue and that a service representative would replace the misassembled photometer.
- Root cause
- Employee error
- Status
- Terminated
- Product code
- JJE
- Quantity affected
- 15
- Distribution
- FL LA AR IA OH OH WI IL LA KS MT
- Date initiated
- 2014-05-27
- Date posted
- 2014-07-30
- Date terminated
- 2014-12-15
- Location
- Newark, DE