Atlas FDA

Dimension Integrated Chemistry Systems intended to measure a variety of analytes in human body fluids.

FDA device recall Z-2133-2014 · posted 2014-07-30 · Terminated

Recall details

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Reason for recall
There is a misassembly issue with the photometer installed on Dimension Integrated Chemistry Systems instruments. Through internal investigation by Siemens Healthcare Diagnostics has confirmed that a limited number of Dimension photometers were misassembled. These photometers are performing within specifications, but may lead to a shortened life of the source lamp.
Action taken
An urgent field safety notice, dated May 2014, was sent to users to inform them of the issue and that a service representative would replace the misassembled photometer.
Root cause
Employee error
Status
Terminated
Product code
JJE
Quantity affected
15
Distribution
FL LA AR IA OH OH WI IL LA KS MT
Date initiated
2014-05-27
Date posted
2014-07-30
Date terminated
2014-12-15
Location
Newark, DE