Dimension EXL LOCI Module TSHL Flex Reagent Cartridge (REF RF612). Intended to quantitatively measure human thyroid stim

FDA device recall Z-1883-2013 · posted 2013-08-06 · Terminated

Recall details

Recalling firm
Siemens Healthcare Diagnostics, Inc
Reason for recall
A rare variant of TSH, identified in a small cluster of patients, is not detected.
Action taken
On May 10, 2013, an Important Safety Information notice was sent to all affected Siemens Healthcare Diagnostic customers in the United States. On May 10, 2013, an Urgent Field Safety Notice was sent to all customers outside of the United States. These notices inform the customer that there is a rare variant of TSH, identified in a small cluster of patients that is not detected by some Siemens assays. This notice informs customers that they should continue to use the affected TSH assays as the observed rate of occurrence during a 30-month time period was 0.6 X 10 -7 and there are strong mitigating factors. These mitigating factors include co-interpretation of TSH results in conjunction with thyroxine and TS, patient history and clinical signs and symptoms. Although Siemens believes a look back is not required as the expected frequency of the variant is extremely rare, these notices state that this decision should be made by the medical director.
Root cause
Device Design
Status
Terminated
Product code
JLW
Quantity affected
33506
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Canada, AE, AR, AT, AU, BA, BD, BE, BG, BO, BR, BY, CH, CL, CM, CN, CO, CR, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FR, GB, GR, GT, HK,
Date initiated
2013-05-10
Date posted
2013-08-06
Date terminated
2016-05-10
Location
East Walpole, MA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
DIMENSION TSHL FLEX REAGENT CARTRIDGE AND TSH SAMPLE DILUENT WITH MODELS RF 612 AND KD691 (K081074)Siemens Healthcare Diagnostics, Inc.2008-05-23