Atlas FDA

Digital VIPR ROC-9839 Roberts Oxygen Ready to Use Digital VIPR (head only) ROC-9840, Roberts Oxygen Ready to Use Digital

FDA device recall Z-1376-2023 · posted 2023-04-12 · Open, Classified

Recall details

Recalling firm
Western/Scott Fetzer Company
Reason for recall
Gauge malfunction: Digital gauge displays an incorrect tank pressure - displays pressure higher than actual tank pressure. This may lead to tank running out of oxygen during use without audible alarms, potentially leading to an out of oxygen event and no flow.
Action taken
On January 3, 2023, the firm issued a Safety Alert to customers. Customers were provided with updated operating instructions. Customers instructed to take the following immediate actions: 1. Implement updated filling and venting instructions which include steps for the filler to verify proper gauge pressure after vent down and after fill/re-fill. 2. Implement an additional verification of gauge pressure upon delivery at the healthcare facility. 3. Take care to minimize any impact to the cylinder during use, transportation, or processing during fill/re-fill. The firm continues to evaluate the gauge error issue and will implement additional corrective actions as appropriate.
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
CAN
Quantity affected
24,287 units
Distribution
US Nationwide distribution in the states of CO, CT, ID, MD, MT, TN, WA.
Date initiated
2023-01-03
Date posted
2023-04-12
Location
Westlake, OH

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.