Atlas FDA

Digital Temple Thermometer labeled under: Bestmed, Good Neighbor (Amerisource Bergen), Kroger, Medline, Meijer, Premier

FDA device recall Z-0412-2016 · posted 2015-12-16 · Terminated

Recall details

Recalling firm
Bestmed, LLC
Reason for recall
An incorrect calibration of the thermometer that causes the device to display inaccurate and frequently lower temperatures than the actual body temperature of the user.
Action taken
Consignees were notified verbally about 11/12/15. Written notification is planed and will include instructions on the identification of the Affected Device, potential health risk, method for return, and manner of restitution for the cost of the Affected Device.
Root cause
Process control
Status
Terminated
Product code
FLL
Quantity affected
305,548
Distribution
US Nationwide Distribution and the country of Canada.
Date initiated
2015-11-12
Date posted
2015-12-16
Date terminated
2016-11-30
Location
Golden, CO

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
DIGITAL FOREHEAD THERMOMETER (K103617)K-Jump Health Co., Ltd.2011-04-04