Atlas FDA

DiaSorin ETI-EA-G assay kit Catalog No. 7570, lot # 113963C

FDA device recall Z-0658-05 · posted 2005-03-29 · Terminated

Recall details

Recalling firm
Diasorin Inc.
Reason for recall
The DiaSorin ETI-EA-G kits contained the wrong microtiter plate. The plate provided in the kits was for a different product, the ProTrac II Tacrolimus kit.
Action taken
Customers were notified of the product and problem via telephone and instructed to check any remaining inventory and discard all incorrect plates. Replacement plates were offered.
Root cause
Other
Status
Terminated
Product code
LSE
Quantity affected
46
Distribution
FL, TX & UT
Date initiated
2004-11-01
Date posted
2005-03-29
Date terminated
2006-07-22
Location
Stillwater, MN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
ETI-EA-G ASSAY (K992191)DiaSorin, Inc.1999-07-12