DEXLOCK Achilles Repair Implant Kits, MAKT4520

FDA device recall Z-0911-2026 · posted 2025-12-11 · Open, Classified

Recall details

Recalling firm
Medline Industries, LP
Reason for recall
There have been multiple complaints of drill bit fusing to the bushing/sleeve within the drill guide during use. Retrieving a replacement drill bit and drill guide may prolong the surgical procedure.
Action taken
Firm began notifying consignees on 11/11/2025 via Urgent Medical Device Recall letters. Customers have been instructed to over-label kits with a sticker instructing them to not use the drill guide within the Achilles Kit and to use the reusable 3.6mm drill guide from the Soft Tissue Instrument Tray. Customers who do not have a Soft Tissue Instrument Tray have been instructed to notify Medline so that a tray can be provided. Customers were instructed to quarantine all affected product in stock and return a completed response form to Medline. Upon receipt of the response form, the firm will then provide over-labels to place on affected inventory. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
NWN
Quantity affected
163 US
Distribution
US Nationwide distribution in the states of AL, AR, AZ, CO, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, MT, NC, NJ, OH, OK, PA, SC, TX, VA, WI.
Date initiated
2025-11-11
Date posted
2025-12-11
Location
Northfield, IL