DeRoyal SafeLiner Suction Canister, 1000cc, REF PHESL-1000B

FDA device recall Z-1283-2023 · posted 2023-03-23 · Open, Classified

Recall details

Recalling firm
DeRoyal Industries Inc
Reason for recall
The reason for the voluntary recall is due to shrinkage of the canister lid. This shrinkage causes the lid not to contact the outer shell to create a seal and thus, it does not create and hold vacuum.
Action taken
The firm issued an URGENT DEROYAL RECALL NOTICE its consignees by email and letter beginning on 02/09/2023. The notice explained the problem and risk, and requested the following actions be taken: Identify and quarantine any affected product found in inventory. If the product has been further distributed, please notify then of this recall. Firm is seeking product return. Questions about receiving credit for return merchandise, contact Jennifer Marsh at jmarsh@deroyal.com or by phone 865-362-4203 or email us at recalls@deroyal.com. If you have questions regarding this recall please contact us at 865-362-6465 or email us at recalls@deroyal.com.
Root cause
Process control
Status
Open, Classified
Product code
KDQ
Quantity affected
27,200 devices
Distribution
US Nationwide distribution in the states of Arizona, Texas, Virginia, Pennsylvania, California, Missouri, North Carolina, Illinois, Indiana, Ohio, Arkansas.
Date initiated
2023-02-09
Date posted
2023-03-23
Location
Powell, TN