DeRoyal KNEE ARTHROSCOPY PACK, REF 89-6192.04, 1 Per Pack, Distributed by: DeRoyal Industries, Inc.; Manufacturer: DeRoy
FDA device recall Z-2124-2014 · posted 2014-07-29 · Terminated
Recall details
- Recalling firm
- DeRoyal Industries Inc
- Reason for recall
- The firm distributed surgical kits which contained Irrigation Sets which were subsequently recalled by Hospira.
- Action taken
- The recall was initiated by letter delivered via UPS on 6/4/2014. The firm requested that the consignee contact Stericycle to arrange for return and replacement of the product. A second notice was sent on 7/22/2014 to those accounts who did not respond to the initial notice. Distributors were requested to notify their end-users or provide a list of end-users to DeRoyal for direct notification. Effectiveness checks will be sent via email for those customers who have provided an email contact and via UPS for those customers who have not responded.
- Root cause
- Packaging change control
- Status
- Terminated
- Product code
- LRO
- Quantity affected
- 128 units
- Distribution
- Distributed in the states of MI, GA, OH, IL, MO, and IN.
- Date initiated
- 2014-06-04
- Date posted
- 2014-07-29
- Date terminated
- 2015-05-05
- Location
- Powell, TN