Depuy Mitek Ratchet Handle with Quick Connect Product code: 219970. Instrument used with anchor inserters for insertion

FDA device recall Z-3106-2011 · posted 2011-08-31 · Terminated

Recall details

Recalling firm
DePuy Mitek, Inc., a Johnson & Johnson Co.
Reason for recall
Quick Disconnect Ratchet Handle may slip up the driver shaft or disengage leaving the shaft in the screw head.
Action taken
Depuy Mitek issued Original Customer Notification Letter on 6/27/11 and an Updated letter to clarify the lot number information sent 6/29/11. Hospitals requested to immediately check all inventories for affected product and return for repalcement product. Also notify surgeons aware of this potential issue and potential risk of a delay in surgery. questions or concerns with regard to this program, please contact your local DePuy Mitek Representative or DePuy Mitek Customer Service directly at 1-800-382-4682, PROMPT 3.
Root cause
Process change control
Status
Terminated
Product code
LXH
Quantity affected
75 units
Distribution
Nationwide distribution.
Date initiated
2011-06-27
Date posted
2011-08-31
Date terminated
2012-01-19
Location
Raynham, MA