Dental chairs distributed under several brand names such as DCI Equipment; Marus MaxStar; Marus NuStar; and Pelton & Cra
FDA device recall Z-1958-2011 · posted 2011-04-12 · Terminated
Recall details
- Recalling firm
- Dental Equipment LLC
- Reason for recall
- Link arm assembly that mounts the rear mounted accessories to the dental chair may potentially break and fall off the dental chair.
- Action taken
- On 03/08/11, Dental Equipment LLC. started sending out the Medical Device Recall notification letter via USPS 1st class mail to US consignees and Federal Express mail to foreign consignees. The firm also sent a recall notification letter to consignees who only purchased dental accessories such as Rear Mounted Cuspidor with Coverset, Arm Articulating Rear Mounted Cuspidor - Marus Gray, Maxstar Orbit Cuspidor utility W/ Porcelain bowl, and zinc plated bolts. The letter identified the affected dental chair as Marus brand with the following model numbers: DC1490, DC1690, DC1700 and DC1702. The firm's corrective action including the replacement of the affected link arm assembly with a new link arm assembly. The firm's field representative will contact each consignee for an appointment to replace the affected link arm assembly. Consignees can contact the Dental Equipment LLC at 503-537-3617 for any questions.
- Root cause
- Process control
- Status
- Terminated
- Product code
- NRU
- Quantity affected
- 55 chairs and 11 replacement bolts distributed to US and foreign consignees outside the US.
- Distribution
- Worldwide distribution to USA, Australia, Canada, Ireland, South Africa, and Vietnam.
- Date initiated
- 2011-03-01
- Date posted
- 2011-04-12
- Date terminated
- 2011-09-21
- Location
- Newberg, OR