Deltec Cozmo Insulin Pump, Model 1700, Smiths Medical International Ltd WD24 4LG UK, Smiths Medical MD, Inc., St Paul, M

FDA device recall Z-1622-2008 · posted 2008-08-15 · Terminated

Recall details

Recalling firm
Smiths Medical MD, Inc.
Reason for recall
Smith Medical discovered an issue with a motor component in specific serial numbers of the Deltec Cozmo Insulin Pump, Model Numbers 1700 and 1800. This issue affects how the motor operates and may cause an over-delivery of insulin which could result in injury to the user.
Action taken
Consignees were mailed an Urgent Medical Device Recall letter dated March 10, 2008. The Urgent letter addressed to "Distributors, Clinicans, Patients and Other Customers" contains Description of the Problem, Results of Investigation, and Advise on Action to be taken by the User. The patient users will be required to return a Recall Confrirmation Form. The USA distributors, USA prescribing physicians and the USA managed Care Accounts (Insurance Companies) will be mailed also a cover letter.
Root cause
Component change control
Status
Terminated
Product code
LZG
Quantity affected
436
Distribution
Class II Recall - Worldwide Distribution --- USA including states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME. RO, MN, MO, MS, ME, NC, ND, NE, NH, NJ, NM, NV, NY,
Date initiated
2008-03-10
Date posted
2008-08-15
Date terminated
2011-01-13
Location
Saint Paul, MN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
DELTEC COZMO INSULIN INFUSION PUMP WITH COZMONITOR BLOOD GLUCOSE METER, MODELS 21-1815, 21-1816, 21-1817 (K062323)Smiths Medical MD, Inc.2006-11-01
SMITHS MEDICAL INSULIN INFUSION PUMP WITH COZMONITOR GLUCOSE METER (K031738)Smiths Medical MD, Inc.2004-05-27
DELTEC INSULIN INFUSION PUMP AND ACCESSORIES (K020655)Deltec, Inc.2002-08-13