Deltec Cozmo Insulin Pump, Model 1700, Product Catalog Numbers 21-1711-01, 21-1712-01 and 21-1713-01. The 3 versions var

FDA device recall Z-0798-03 · posted 2003-05-09 · Terminated

Recall details

Recalling firm
Deltec, Inc
Reason for recall
A software error could have resulted in the delivery of a larger bolus of insulin than anticipated when the pump was programmed in a particular manner.
Action taken
Recall Notification letters were sent to the consignees on March 14, 2003. The letters gave recommendations for avoiding the problem, stated that updated software to prevent the problem would be introduced, and requested that the customer call a toll-free phone number to arrange replacement of their affected pump.
Root cause
Other
Status
Terminated
Product code
LZG
Quantity affected
913 pumps
Distribution
The product was distributed nationwide in the United States. There are no foreign or U.S. Government accounts.
Date initiated
2003-03-14
Date posted
2003-05-09
Date terminated
2004-07-02