Deltec branded CADD-Prizm PCS II ambulatory infusion pump and the CADD-Prizm VIP system ambulatory infusion pumps; (Both

FDA device recall Z-0718-2007 · posted 2007-04-05 · Terminated

Recall details

Recalling firm
Smiths Medical MD, Inc.
Reason for recall
The year of manufacture had been incorrectly entered as 1980 into the pumps software during the manufacture process. As a result the pump may go into an immediate 'Clock battery is Low/Service Immediately' or 'Clock Battery needs service soon' alarm message indicating that the pump requires servicing.
Action taken
All consignees who have received the affected CADD-Prizm PCS II ambulatory infusion pumps and the CADD-Prizm VIP System ambulatory infusion pumps were contacted via telephone by February 1, 2007. The problem will be explained to the consignee and they will be asked to return the pumps to Smiths Medical MD in order for the problem to be corrected.
Root cause
Other
Status
Terminated
Product code
FRN
Quantity affected
27
Distribution
Nationwide.
Date initiated
2007-01-19
Date posted
2007-04-05
Date terminated
2007-05-25
Location
Saint Paul, MN