Defibtech Sentry Semi-Automatic External Defibrilator (AED) sold as 'Lifeline AED' brand -Model DDU-100A (version 2.002
FDA device recall Z-0580-2007 · posted 2007-03-08 · Terminated
Recall details
- Recalling firm
- Defibtech, LLC
- Reason for recall
- Device Malfunction-The self-test software may allow a self-test to clear a previously detected low battery condition. If this occurs, the operator may be unaware of the low battery and the device may be "unable" to deliver a defibrillation shock, which could result in failure to resuscitate a patient.
- Action taken
- Defibtech LLC initiated notification to customers/distributors on 2/17/07 by email. Hard copy letters followed on 2/22/07 with acknowledgement card. Users provided with instructions for 'Immediate Recommendations' until the unit''s software is upgraded. Distributors, their customers, and direct accounts instructed to locate affected units and upgrade these units in the field. Distributors, their customers, and direct accounts shall return to Defibtech documentation confirming unit software upgrade was completed..Press was issued by the firm on March 6, 2007
- Root cause
- Software design
- Status
- Terminated
- Product code
- MKJ
- Quantity affected
- 22,197 units
- Distribution
- Worldwide distribution ---- USA and countries of Canada, Uruguay, Jordon, Turkey United Kingdom, Saudi Arabia, Sweden France, South Africa, Cyprus, Spain, Germany India, Japan, Israel, Australia, Norw
- Date initiated
- 2007-02-17
- Date posted
- 2007-03-08
- Date terminated
- 2021-06-09
- Location
- Guilford, CT