DCTM Digital Ceiling Tube Mount, model DCTM; a fully counter-balanced radiographic x-ray tube suspension system designed

FDA device recall Z-0712-04 · posted 2004-07-20 · Terminated

Recall details

Recalling firm
Del Medical Systems Group
Reason for recall
There is a potentially defective worm gear in the tension adjusting mechanism of the balancer used in the Digital Ceiling Mounted Tube Support System for overhead x-ray units.
Action taken
Del Medical sent a Field Correction Recall letter dated 2/13/04 to all dealers/assemblers receiving DCTM''s during the affected time frame, alerting them to potential defective worm gear. The accounts were asked to locate the systems listed in the attachment to the letter, and alert the end users to take the units out of service until they can be inspected. Follow-up letters were sent via registerd mail on 2/19/04, reiterating the 2/13/04 letter and providing the accounts with instructions for inspection of the balancers. If the worm gear fails the test, the dealers were instructed to contact Del Medical for replacement of the balancer and assure that the unit is not placed back in service until the balancer has been replaced.
Root cause
Other
Status
Terminated
Product code
IYB
Quantity affected
57 units
Distribution
Nevada, Ohio, Tennessee, Nebraska, Arizona, Maryland, Wisconsin, Florida, Texas, Kansas, Oklahoma, Illinois, Colorado, California, Minnesota, Michigan, Alabama, Rhode Island and internationally to Can
Date initiated
2004-02-13
Date posted
2004-07-20
Date terminated
2004-09-30
Location
Franklin Park, IL