DCI Equipment Track Mount Light Assembly, 110v
FDA device recall Z-1077-2007 · posted 2007-07-12 · Terminated
Recall details
- Recalling firm
- Dental Equipment LLC
- Reason for recall
- Potential for light shield to melt due to light assemblies not properly rated to handle the VAC.
- Action taken
- On 5/1/07 the dental office contacted the manufacturer via telephone regarding the incident. On 5/2/07 the firm's service representative visited the dental office and replaced the lights.
- Root cause
- Other
- Status
- Terminated
- Product code
- EAZ
- Quantity affected
- 4 dual track lights
- Distribution
- The lights were distributed to one distributor in NY who in turn distributed the lights to one dental office.
- Date initiated
- 2007-05-01
- Date posted
- 2007-07-12
- Date terminated
- 2007-07-12
- Location
- Newberg, OR