Data reduction program written specifically for Bio-Tek ELx800 automated microtiter plate reader with use of Diastat Ant

FDA device recall Z-0673-05 · posted 2005-04-06 · Terminated

Recall details

Recalling firm
Diasorin Inc.
Reason for recall
The data reduction program written for a specific automated microtiter plate instrument (Bio-Tek ELx800 reader) was found to be faulty. This program is used to calculate the cut off values for Diastat Anti-Mitochondrial Antibody Kit. Program should include a borderline specification to indicate repeat testing is needed.
Action taken
A notification letter and effectiveness check form was sent to consignees. This letter recommended that users manually verify results using the specifications outlined in the package insert until a new software version is installed.
Root cause
Other
Status
Terminated
Product code
DBM
Quantity affected
8
Distribution
CA, CT, KS, MA & WA
Date initiated
2004-04-12
Date posted
2005-04-06
Date terminated
2006-07-22
Location
Stillwater, MN