Atlas FDA

D-RAD Self-Tapping Locking Screw intended for the fixation of fractures involving the distal radius. smith&nephew D-RAD

FDA device recall Z-1317-2018 · posted 2018-03-20 · Terminated

Recall details

Recalling firm
Smith & Nephew, Inc.
Reason for recall
One lot of D-RAD Self-Taping Locking Screws used with the Distal Radius Fracture Kit were machined out of specification. Screws measured over tolerance within the head thread form by 0.002-0.003 inch.
Action taken
The firm initiated their recall by letter and email on 03/07/2018. The firm requested return of the product.
Root cause
Under Investigation by firm
Status
Terminated
Product code
HWC
Quantity affected
59 units
Distribution
US, Puerto Rico, Malaysia
Date initiated
2018-03-07
Date posted
2018-03-20
Date terminated
2018-10-31
Location
Memphis, TN