D 744 Chloride membrane, CI, REF Model #942-061, accessory to the ABL700 blood analyzer.

FDA device recall Z-1314-06 · posted 2006-08-01 · Terminated

Recall details

Recalling firm
Radiometer America Inc
Reason for recall
The CI membrane units generate too low values for status and sensitivity during calibration.
Action taken
The recalling firm sent a letter, dated 2/05, to all of their consignees.The letter instructed the customers to discontinue use of the product and discard any remaining stock which they had on hand. The letter was accompanied by replacement chloride membranes to replace the customer''s portion of the suspect lot.
Root cause
Other
Status
Terminated
Product code
CGZ
Quantity affected
85/4 unit boxes
Distribution
Nationwide.
Date initiated
2005-02-01
Date posted
2006-08-01
Date terminated
2011-11-18
Location
Westlake, OH