CyberKnife Robotic Radiosurgery System, Synchrony Respiratory Tracking System (Standard Treatment Couch). Synchrony Resp

FDA device recall Z-0035-2012 · posted 2011-10-11 · Terminated

Recall details

Recalling firm
Accuray Incorporated
Reason for recall
A report and investigation determined that the set screws, which were part of the retrofit of a previous recall were incorrectly engaged which allowed rotation and subsequent descent of the assembly.
Action taken
Urgent Device Correction Letters were sent to all sites affected by the retrofit on September 12, 2011 via Fed Ex. The letter identified the affected product and the description of the potential problem. Customers were asked to inspect their Synchrony System by following the instructions provided. If the lower extension feels loose, customers should contact Accuray Customer Support immediately to schedule service of the system. The letter also states that all affected systems will be inspected and corrected, if necessary by Accuray to prevent the potential problem. Accuray Customer Support will contact customers to schedule the inspection. Questions or concerns regarding the issue should be directed to Accuray Customer Support at 1-877-8667 for US customers and 1-408-716-4700 for customers outside of the US, or customersupport@accuray.com.
Root cause
Process control
Status
Terminated
Product code
IYE
Quantity affected
236 total devices
Distribution
Worldwide Distribution
Date initiated
2011-09-12
Date posted
2011-10-11
Date terminated
2012-02-27
Location
Sunnyvale, CA