CSA Medical truFreeze System; Model: CC3-01, a cryogenic surgical device ***The involved units are those with the newest

FDA device recall Z-0941-2015 · posted 2015-01-07 · Terminated

Recall details

Recalling firm
CSA Medical
Reason for recall
An increase in complaint trending for a loss of audible sound associated with the timer has been observed.
Action taken
The firm, CSA Medical, sent a letter to its consignees 11/14/2014, via Fedex overnight . The letter explains the event and asks consignees to us a verbal call of time during the cryogen spray period until the root cause is determined and a permanent correction is implemented. Consignees are also asked to complete an acknowledgment form and return it to CSA Medical via, fax, email, telephone, etc.
Root cause
Software design
Status
Terminated
Product code
GEH
Quantity affected
46
Distribution
US Distribution to states of: AR, CA, CO, CT, DC, FL, GA, IL, IN, LA, MA, MD, MI, NC, NJ, NY, OK, PA, RI, SC, TN, TX, UT, VA, and WV.
Date initiated
2014-11-10
Date posted
2015-01-07
Date terminated
2015-03-25
Location
Lexington, MA