CrossCHECK Non-Locking Screw in the following sizes: 3.0 x 8 mm, REF CCP-N3008; 3.0 x 10 mm, REF CCP-N3010; 3.0 x 12 mm,

FDA device recall Z-2542-2015 · posted 2015-09-01 · Terminated

Recall details

Recalling firm
Wright Medical Technology, Inc.
Reason for recall
These screws have a potential of passing through the plate in the case when torque is applied in excessive amounts, which may result in intraoperative malfunction.
Action taken
Wright sent an Urgent Medical Device Recall letter dated July 20, 2015, to all affected customers. Customers were asked to complete the attached response form confirming receipt of the notice and the status of any products on hand, and return a copy to Wright by email (complaints@wmt.com or fax (901)867-7401. Also, if customers have any product in their inventory the product should be sent back to Arlington, TN (ALM). For questions regarding this recall call 901-290-5662.
Root cause
Employee error
Status
Terminated
Product code
HRS
Quantity affected
5687 units
Distribution
Nationwide Distribution
Date initiated
2015-07-21
Date posted
2015-09-01
Date terminated
2018-08-01