CRANIOTOMY PACK, Pack Number DYNJ46552O
FDA device recall Z-3069-2024 · posted 2024-09-06 · Open, Classified
Recall details
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Reason for recall
- A non-sterile Mastisol component, which was not labeled as being sterile or non-sterile, was placed in the non-sterile portion of certain sterile convenience kits and distributed. In the event that the user assumes sterility during its use, there is an increased risk of patient infection.
- Action taken
- Medline Industries issued a Recall notice to its consignees on 06/18/2024 via email. The notice explained the issue, potential risk, and directed the product to be stickered to instruct the used to remove the recalled component from the kit and replace from sterile supply. Distributors or those who have transferred the product were directed to notify those to whom the product was distributed of the recall. For questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
- Root cause
- Process change control
- Status
- Open, Classified
- Product code
- OJH
- Quantity affected
- 188 units
- Distribution
- US: TX, IL, CA
- Date initiated
- 2024-06-18
- Date posted
- 2024-09-06
- Location
- Northfield, IL
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.