Atlas FDA

CPB CATHETER KIT- DR.VAN TRIGHT. The Sarns High Flow Aortic Arch Cannula is indicated for use in perfusion of the ascend

FDA device recall Z-1842-2012 · posted 2012-06-20 · Terminated

Recall details

Recalling firm
Terumo Cardiovascular Systems Corporation
Reason for recall
During production of one lot of the Sarns High-Flow Aortic Arch Cannula, Terumo Cardiovascular Systems (Terumo CVS) found a foreign substance on the outer surface of some cannulae connectors. Terumo CVS' preliminary investigation found that the substance can be dislodged from the cannula surface, the substance was likely deposited during the molding process, but the exact composition of the parti
Action taken
Terumo sent an "URGENT MEDICAL DEVICE RECALL" letter dated May 9, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter advises customers to discontinue use of the affected product and to return the product to the firm. A Response Form was attached for customers to complete and return. Contact Terumo CVS Customer Service at 1-800-521-2818 for questions regarding this recall.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
DWF
Quantity affected
133,793 (each)
Distribution
Worldwide Distribution-USA (nationwide) including the states of AL , AR , AZ, CA , CO , CT , FL , GA , HI , IA , IL , IN , KS , KY , LA , MA , MD , ME , MI , MI , MN , MO , MT , NC , ND , NE , NH , NJ
Date initiated
2012-05-09
Date posted
2012-06-20
Date terminated
2013-07-11
Location
Ann Arbor, MI