Covidien Kangaroo Polyurethane Feeding Tube Radiopaque Line, Safe Enteral Connections 6.5 Fr/Ch (2.2 mm) x 36" (91 cm),

FDA device recall Z-1033-2016 · posted 2016-03-02 · Terminated

Recall details

Recalling firm
Covidien LLC
Reason for recall
8FR 16 PVC tubes were packaged in a 6.5 FR 36 Polyurethane packs and 8FR 20 PVC tubes were packaged in an 8FR 20 Polyurethane packs
Action taken
Medtronic/Covidien Customers notified via an Urgent Medical Device Recall letter, dated January 22, 2016. The letter identifies the affected product, instructs customers to discontinue use and quarantine any of the product, and return all affected inventory. Customers are also to complete the "Recalled Product Return Form." Questions or concerns can be directed to a Medtronic representative or Customer Service at 800-882-5878.
Root cause
Labeling Change Control
Status
Terminated
Product code
KNT
Quantity affected
1,010 units
Distribution
Worldwide Distribution -- US, to the state of CA; and, to Australia and Europe.
Date initiated
2016-01-21
Date posted
2016-03-02
Date terminated
2017-06-01
Location
North Haven, CT