CoverStainer Slide Rack The CoverStainer Slide Rack is an accessory to the CoverStainer instrument. The purpose is to ho
FDA device recall Z-0563-2013 · posted 2012-12-19 · Terminated
Recall details
- Recalling firm
- Dako Denmark A/S
- Reason for recall
- The expected life time of the CoverStainer slide racks (CS 10330) does not meet established performance requirements, originating from variations in the manufacturing process.
- Action taken
- Dako sent a Voluntary Product Recall Notification dated November 13, 2012, to all affected customers. This voluntary recall notification was initiated because the expected life time of the CoverStainer slide racks (CS 10330) does not meet established performance requirements, originating from variations in the manufacturing process. The result can be that slides are falling out of the rack during the staining process. While this has been observed at only a few customer sites, the corrective action includes all CoverStainer installations. Customers were notified that Dako would correct the rack situation by using a new manufacturer and has imroved the rack design. Customers with questions were instructed to contact their local Dako representative. For questions regarding this recall call 805-566-3036.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- KPA
- Quantity affected
- 25 Units US (50-100 racks per unit) 106 total
- Distribution
- Worldwide Distribution - USA (nationwide) and the following countries: Australia France Rep. of Korea Belgium Germany Russia Brazil Israel Saudi Arabia Canada Italy Spain Chile Japan Sweden China Mexi
- Date initiated
- 2012-11-26
- Date posted
- 2012-12-19
- Date terminated
- 2013-05-14
- Location
- Glostrup, Denmark