COULTER PrepPlus, Part Number: 286600 Product Usage: The COULTER PrepPlus, is a microprocessor-controlled pipetting and
FDA device recall Z-0002-2013 · posted 2012-10-01 · Terminated
Recall details
- Recalling firm
- Beckman Coulter Inc.
- Reason for recall
- The recall was initiated because Beckman Coulter has confirmed that the Coulter PrepPlus may experience intermittant reagent level sense failures where the instrument may not produce a warning that a reagent vial is low or out of reagent.
- Action taken
- Beckman Coulter sent an Urgent Product Correction letter dated January 20, 2012 to all their customers who purchased the affected product. The letter provides the customers with an explanation of the problem identified and actions to be taken. Customers were instructed to complete and return the Response Form with in 10 days. Customers with questions regarding the PCA letter were instructed to contact Beckman Coulter Service at (800) 526-7694 in the US and Canada or contact their local Beckman Coulter Representative.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- JQW
- Quantity affected
- 36 units total (15 units in the US)
- Distribution
- Worldwide Distribution - US Nationwide and the countries of: Australia, Botswana, Canada, France, Germany, Japan, Malawi and Sweden.
- Date initiated
- 2012-01-20
- Date posted
- 2012-10-01
- Date terminated
- 2012-10-02
- Location
- Brea, CA