Coulter LH 780 Hematology Analyzer, Part No. 723585, A90728 Product Usage: Coulter LH 780 Hematology Analyzer is a is a
FDA device recall Z-0471-2014 · posted 2013-12-10 · Terminated
Recall details
- Recalling firm
- Beckman Coulter Inc.
- Reason for recall
- Beckman Coulter is recalling the LH 750 and LH 780 Hematology Systems because the optical degradation may potentially cause a delay in reporting results for nucleated red blood cells (nRBC), white blood cell differential and/or reticulocytes.
- Action taken
- Beckman Coulter sent an customer notification letter dated 11/8/13 to all customers who purchased the LH 750 and LH 780 Hematology Systems. The letter informs the customers of the problems identified and the actions to be taken. Customers are instructed to complete and return the enclosed response form within 10 days. Customers with questions are instructed to contact Beckman Coulter's Customer Support Center at http://www.beckmancoulter.com/customer/support or call (800) 526-7694 in the US and Canada. Outside the US and Canada, customers are instructed to contact Beckman Coulter Representative. On 2/17/14, Beckman Coulter expanded the recall to inform Coulter LH 750 and Coulter LH 780 Hematology Analyzer customers of the issue related to LH Cleaner and Coulter Clenz.
- Root cause
- Under Investigation by firm
- Status
- Terminated
- Product code
- GKZ
- Quantity affected
- 1961 units total (610 units in US)
- Distribution
- Worldwide Distribtion - USA Nationwide including the countries of Algeria, Andorra, Angola, Australia, Austria, Bahrain, Bangladesh, Belgium, Bermuda, Bosnia and Herzegovina, Bulgaria, China, Colombia
- Date initiated
- 2013-11-08
- Date posted
- 2013-12-10
- Date terminated
- 2014-12-25
- Location
- Brea, CA