Atlas FDA

COULTER DxH Diluent, Catalog No. 628017, Part No. A59956. For use as an isotonic buffered diluent in conjunction with a

FDA device recall Z-1616-2015 · posted 2015-05-15 · Terminated

Recall details

Recalling firm
Beckman Coulter Inc.
Reason for recall
Beckman Coulter has received an increased number of complaints related to Hemoglobin (HGB) on UniCel DxH 800 and DxH 600 Coulter Cellular Analysis Systems. The use of the DxH Diluent lots listed in the recall notice may eventually result in a compromised HGB chamber.
Action taken
An Urgent Medical Device Recall letter dated 4/13/15 was sent to customers who purchased the Coulter DxH Diluent. The letter informs customers that Beckman Coulter has received an increased number of complaints related to Hemoglobin (HGB) on UniCel DxH 800 and DxH 600 Coulter Cellular Analysis Systems. The letter informs customers of the problems identified and the actions to be taken. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with questions are instructed to contact Customer Support Center via http://www.beckmancoulter.com/customersupport/support or call (800) 526-7694 in the US and Canada. Customers outside the US and Canada are instructed to contact their local Beckman Coulter Representative.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
GIF
Quantity affected
829,888 units total (659,584 units in US)
Distribution
Worldwide Distribution -- Argentina; Brazil; Brunei Darussalam; Canada; Chile; Colombia; Hong Kong; India; Lebanon; Macao; Mexico; Panama; Philippines; Qatar; Singapore; Taiwan; Thailand; Trinidad and
Date initiated
2015-04-13
Date posted
2015-05-15
Date terminated
2015-10-01
Location
Brea, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM (K120771)Beckman Coulter, Inc.2013-03-22
UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM (K081930)Beckman Coulter, Inc.2008-12-19