COULTER DxH Diluent, Catalog No. 628017, Part No. A59956. For use as an isotonic buffered diluent in conjunction with a
FDA device recall Z-1616-2015 · posted 2015-05-15 · Terminated
Recall details
- Recalling firm
- Beckman Coulter Inc.
- Reason for recall
- Beckman Coulter has received an increased number of complaints related to Hemoglobin (HGB) on UniCel DxH 800 and DxH 600 Coulter Cellular Analysis Systems. The use of the DxH Diluent lots listed in the recall notice may eventually result in a compromised HGB chamber.
- Action taken
- An Urgent Medical Device Recall letter dated 4/13/15 was sent to customers who purchased the Coulter DxH Diluent. The letter informs customers that Beckman Coulter has received an increased number of complaints related to Hemoglobin (HGB) on UniCel DxH 800 and DxH 600 Coulter Cellular Analysis Systems. The letter informs customers of the problems identified and the actions to be taken. Customers are instructed to complete and return the enclosed Response Form within 10 days. Customers with questions are instructed to contact Customer Support Center via http://www.beckmancoulter.com/customersupport/support or call (800) 526-7694 in the US and Canada. Customers outside the US and Canada are instructed to contact their local Beckman Coulter Representative.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- GIF
- Quantity affected
- 829,888 units total (659,584 units in US)
- Distribution
- Worldwide Distribution -- Argentina; Brazil; Brunei Darussalam; Canada; Chile; Colombia; Hong Kong; India; Lebanon; Macao; Mexico; Panama; Philippines; Qatar; Singapore; Taiwan; Thailand; Trinidad and
- Date initiated
- 2015-04-13
- Date posted
- 2015-05-15
- Date terminated
- 2015-10-01
- Location
- Brea, CA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM (K120771) | Beckman Coulter, Inc. | 2013-03-22 |
| UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM (K081930) | Beckman Coulter, Inc. | 2008-12-19 |