Cotton-Leung Biliary Stent. CLSO-10-10, Stent Length:10cm in length, Disposable-Single use only, Rx Only, Wilson Cook Me
FDA device recall Z-2332-2008 · posted 2008-09-23 · Terminated
Recall details
- Recalling firm
- Cook Endoscopy
- Reason for recall
- It is possible the stents inside CLSO-10-10, lot W2512492 were 5cm in length, instead of 10cm in length as listed on the product packaging label.
- Action taken
- Consignees were notified on/about 07/11/2008 via an Urgent - Product Recall Notice. All medical user consignees were instructed to review the attached list and quarantine any affected product in their possession. They are to return any affected product to Cook Endoscopy. A reply form was attached to verify receipt of notification and quantity product returned. All product returned as part of this recall initiative will be dispositioned to scrap and destroyed.
- Root cause
- Process control
- Status
- Terminated
- Product code
- FGE
- Quantity affected
- 20 units
- Distribution
- Worldwide Distribution --- USA including states of MO, NC, NV, TX, and UT and country of India.
- Date initiated
- 2008-07-02
- Date posted
- 2008-09-23
- Date terminated
- 2009-06-04
- Location
- Winston Salem, NC