Atlas FDA

Cotton-Leung Biliary Stent. CLSO-10-10, Stent Length:10cm in length, Disposable-Single use only, Rx Only, Wilson Cook Me

FDA device recall Z-2332-2008 · posted 2008-09-23 · Terminated

Recall details

Recalling firm
Cook Endoscopy
Reason for recall
It is possible the stents inside CLSO-10-10, lot W2512492 were 5cm in length, instead of 10cm in length as listed on the product packaging label.
Action taken
Consignees were notified on/about 07/11/2008 via an Urgent - Product Recall Notice. All medical user consignees were instructed to review the attached list and quarantine any affected product in their possession. They are to return any affected product to Cook Endoscopy. A reply form was attached to verify receipt of notification and quantity product returned. All product returned as part of this recall initiative will be dispositioned to scrap and destroyed.
Root cause
Process control
Status
Terminated
Product code
FGE
Quantity affected
20 units
Distribution
Worldwide Distribution --- USA including states of MO, NC, NV, TX, and UT and country of India.
Date initiated
2008-07-02
Date posted
2008-09-23
Date terminated
2009-06-04
Location
Winston Salem, NC

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
ZIMMON BILIARY STENT SET (K851962)Wilson-Cook Medical, Inc.1985-12-17