Corin BIOLOX DELTA MOD HEAD,032mm, EX LONG +7mm, REF 104.3215

FDA device recall Z-0726-2023 · posted 2022-12-19 · Open, Classified

Recall details

Recalling firm
Corin Ltd
Reason for recall
The BIOLOX Delta Ceramic Head size 36XL is labelled as the size 32XL and vice-versa.
Action taken
The firm issued a FIELD SAFETY NOTICE/RECALL LETTER on 11/08/2022 by email. The notice explained the issue, potential risk to health, and post implantation recommendations. The following actions were requested: - Return the device from the filed to Coring USA Limited, Paramount Drive, Raynham, MA 02767, USA - Complete the acknowledgment of receipt and forward to vigilance@coringroup.com to confirm receipt of the Field Safety Notice/Recall Letter. For further information concerning this event, contact: vigilance@coringroup.com
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
LZO
Quantity affected
5 units
Distribution
US Nationwide distribution in the states of FL, GA, MA, MI, Ok, PA, TX.
Date initiated
2022-11-08
Date posted
2022-12-19
Location
Cirencester, United Kingdom