CORI REAL INTELLIGENCE Robotic Drill, Product No. ROB10013 and ROB10013S

FDA device recall Z-0708-2024 · posted 2024-01-11 · Open, Classified

Recall details

Recalling firm
Blue Belt Technologies, Inc
Reason for recall
Possibility that the user is unable to resolve drill disconnection error messages.
Action taken
The recalling firm sent an Urgent Medical Device Recall/Correction notice, dated 11/13/23, by mail to consignees. The notice informed consignees that they may continue to use affected devices as is provided that they ensure the backup Robotics Instrument Tray is available during surgery and to review the CORI Surgical System User Manual, which provides instructions for solving error messages. If consignees are unable to resolve the error messages, they should be prepared to convert to manual surgical procedure following the instructions in Appendix D: Recovery Procedure Guidelines of the CORI Surgical System User Manual. In accordance with routine servicing, consignees should return the CORI drill after 75 uses for servicing, during which the recalling firm will make the hardware update required to correct this recall event. Consignees are asked to contact their Account Representative to report pervasive drill disconnection error messages. Completed acknowledgement forms are to be returned to the recalling firm by email at fieldactions@Smith-Nephew.com. Questions are to be directed by email to fieldactions@smith-nephew.com or to your Account Representative.
Root cause
Device Design
Status
Open, Classified
Product code
OLO
Quantity affected
1859 units
Distribution
Domestic: Nationwide Distribution.
Date initiated
2023-11-13
Date posted
2024-01-11
Location
Pittsburgh, PA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
REAL INTELLIGENCE™ CORI™ (K231963)Blue Belt Technologies, Inc.2023-08-01