Atlas FDA

Copper NiTi Preformed Archwires; Copper NiTi 40 BAF .016X.022 LO Pk10, Part No. 211-0952. Intended to provide tooth move

FDA device recall Z-0343-2016 · posted 2015-11-30 · Terminated

Recall details

Recalling firm
Ormco/Sybronendo
Reason for recall
Ormco Corporation is recalling Copper NiTi Preformed Archwires because the product transformation temperature as listed on the labeling is incorrect.
Action taken
A recall letter dated 11/5/15 was sent to all their customers who purchased the Copper NiTi Preformed Archwires because it has been brought to Ormco's attention that the product transformation temperature as listed on the labeling is incorrect. The wire is labeled as having 40 degrees Celsius transformation temperature, but the actual wire has a transformation temperature of 29 degrees Celsius. The letter provides the customers with the problems identified and the actions to be taken. Customers are instructed to complete and fax back the enclosed acknowledgement form.
Root cause
Employee error
Status
Terminated
Product code
DZC
Quantity affected
161 units
Distribution
Distributed US (nationwide) and the countries of Canada, Japan, Mexico, Russia, and Peru.
Date initiated
2015-11-04
Date posted
2015-11-30
Date terminated
2016-01-19
Location
Glendora, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.