Cook Medical Tao Brush I.U.M.C. Endometrial Cytology/Histology Sampler (Tao Brush), 9.0Fr/26 cm, 3.5 cm, sterile, Cook O

FDA device recall Z-1416-2009 · posted 2009-06-16 · Terminated

Recall details

Recalling firm
Cook OB/Gyn, Inc.
Reason for recall
There is an ink on the brush which has not been tested for use with the Tao Brush.
Action taken
The consignee was notified by an Urgent: Recall Notice letter dated 4/2/09. The customer is to forward the notice to all personnel that need to be aware of it; pull from their stock and isolate any affected product; and complete the form. Upon receipt of the form, Cook Women's Health will issue a Return Goods Authorization number and the customer should return the affected products via FedEx using the Return Goods Authorization number.
Root cause
Process control
Status
Terminated
Product code
HFE
Quantity affected
15,396
Distribution
Nationwide Distribution.
Date initiated
2009-04-02
Date posted
2009-06-16
Date terminated
2009-12-03
Location
Spencer, IN