Atlas FDA

Cook Medical Ciaglia Blue Rhino G2 Advanced Percutaneous Tracheostomy Introducer Set, C-PTISY-100-HC-PERC8, G12565.

FDA device recall Z-2083-2010 · posted 2010-07-26 · Terminated

Recall details

Recalling firm
Cook, Inc.
Reason for recall
These set and/or tray products are being recalled because they include a Covidien Shiley 6PERC or Covidien Shiley 8PERC tracheostomy tube component that has been associated with cuff leakage as a result of the pilot balloon inflation assembly.
Action taken
Cook Medical sent an Urgent Product Recall letter dated April 21, 2010, to hospitals and other user facilities receiving product direct from Cook as well as distribution centers receiving direct shipments from Cook. Consignees were directed to cease distribution of product and notify their customers immediately.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
JOH
Quantity affected
28819 total, all products
Distribution
Worldwide distribution: USA, Canada, Mexico, Australia, Chile, Costa Rica, Denmark, Dominican Republic, Guatemala, India, Malaysia, Panama, Peru, Puerto Rico, Republic of Korea, Trinidad and Tobago, a
Date initiated
2010-04-21
Date posted
2010-07-26
Date terminated
2012-01-23
Location
Bloomington, IN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
CIAGLIA BLUE DOLPHIN BALLOON PERCUTANEOUS TRACHEOSTOMY INTRODUCER, MODEL: MULTIPLE, PREFIX C-PTBS (K093469)Cook, Inc.2010-01-08