Convenience kits containing Tego Connectors: Medline DIALYSIS DRESSING CHANGE SKU EBSI1498 Medline DIALYSIS DRESSING CHA

FDA device recall Z-1467-2026 · posted 2026-02-24 · Open, Classified

Recall details

Recalling firm
Medline Industries, LP
Reason for recall
Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.
Action taken
On January 8, 2026, the firm notified affected customers of the product issue through email and first class mail. Customers were instructed to request stickers to over-label affected convenience kits with a warning label indicating to remove the affected component and replace it with product from supply. No kits are to returned.
Root cause
Process control
Status
Open, Classified
Product code
OXQ
Quantity affected
31,848
Distribution
US Nationwide distribution.
Date initiated
2026-01-08
Date posted
2026-02-24
Location
Northfield, IL