Atlas FDA

Convenience kits containing Medline Namic Angiographic Control Syringes with Rotating Adaptor CORONARY ANGIO PACK Model

FDA device recall Z-1981-2026 · posted 2026-05-06 · Open, Classified

Recall details

Recalling firm
American Contract Systems Inc
Reason for recall
Impacted kits contain syringes recalled by Medline Industries. Syringe rotating adaptor may unwind during use, which may result in a loose connection or full disconnection between the syringe and manifold. Potential risk for biohazard exposure, blood loss, infection, or air embolism.
Action taken
Firm notified distributors with an "URGENT: MEDICAL DEVICE RECALL NOTIFICATION" dated March 19, 2026, via email. Distributors were instructed to notify end-users. An updated letter with additional kits was sent to the customer on 3/20/26. Customers should quarantine all affected product. Customers may add warning labels to affected ACS kits to ensure all users are aware of prescribed actions that were identified in the Medline recall notice. The warning labels state, "Per Medline, due to the risk of serious injury or death, customers are directed to remove and destroy all Namic RA Syringes. If use is unavoidable because failure to proceed would result in patient harm, the Namic RA syringe must be used with extreme caution and vigilance, including manual stabilization of syringe-to-manifold connections and continuous, 100% monitoring during setup and throughout the entire procedure." Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. If you have any questions, contact the Regulatory Affairs Manager, by email: gm-omra_recalls@owens-minor.com or at 419-455-2090.
Root cause
Nonconforming Material/Component
Status
Open, Classified
Product code
LRO
Quantity affected
7,311 kits
Distribution
US distribution to MN. No OUS distribution.
Date initiated
2026-03-19
Date posted
2026-05-06
Location
Tiffin, OH

Monitor recalls programmatically