Atlas FDA

Contour Next Control Solution for Contour Next Blood Glucose Test System. Liquid quality control solution. Plastic bottl

FDA device recall Z-2482-2015 · posted 2015-08-29 · Terminated

Recall details

Recalling firm
Bayer Healthcare, LLC
Reason for recall
If the control solution bottle is not thoroughly shaken prior to use, the control test result may be higher than the control range and may not be recognized and marked as a control test by the blood glucose meter. Also, the control solution, used as a quality control test in blood glucose meter systems, may read outside the specified range too high or too low.
Action taken
On 6/22/2015, Bayer Healthcare began sending URGENT - MEDICAL DEVICE FIELD CORRECTIVE ACTION notifications to affected Health Care Providers, Pharmacies, and Users. The recall notification included a description of the reason for the recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. Customers are instructed to always check the expiration, not to use and discard of expired control solutions. Also, only CONTOUR NEXT control solution should be used with the CONTOUR NEXT test strips. The control solution bottle should be shaken about 15 times before each use. The letter instructed customers to refer to their meter's user guide for further testing instructions. If test results outside the control range are received, another test should be performed following the instructions provided in the letter. If the problem persists, customers should contact Bayer Diabetes Support. Bayer will replace the control solution or provide new control ranges for the test strips, depending on the situation. The firm has implemented improvements to provide more consistent control solution product.
Root cause
Device Design
Status
Terminated
Product code
JJX
Quantity affected
5 million bottles of control solution
Distribution
Worldwide Distribution -- USA, Australia, Austria, Belgium, Canada, Croatia, Denmark, Finland, France, Germany, Greece, Ireland, Italy, Luxembourg, Netherlands, Norway, Portugal, Slovenia, Spain, Swed
Date initiated
2015-06-22
Date posted
2015-08-29
Date terminated
2016-05-23
Location
Mishawaka, IN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
CONTOUR (R) NEXT BLOOD GLUCOSE MONITORING SYSTEM (K111268)Bayer Healthcare2012-03-06