CONTAK RENEWAL 3 CRT-D (model no. H170, H173, H175) and CONTAK RENEWAL 3 HE CRT-D(model no. H177 and H179). Cardiac Resy

FDA device recall Z-0949-05 · posted 2005-06-30 · Terminated

Recall details

Recalling firm
Guidant Corporation
Reason for recall
Engineering analysis has determined that the magnetic switch in certain devices may stick in the closed position. When the magnetic switch becomes stuck in the closed position treatment of ventricular or atrial tachyarrhythmias is inhibited, while bradycardia pacing is unaffected.
Action taken
Firm issued a Dr. Doctor letter dated 06/23/05. Letter informs physicians of the issue, clinical implications and recommendations. UPDATE: An Advisory Update, dated 08/01/05 is being issued to physicians to inform them of a new programmer software application upgrade and the programmer screen alerts. Additionally regulatory approval was given for a new switch component that will be incorporated into all Guidant CRT-D products.
Root cause
Other
Status
Terminated
Product code
NIK
Quantity affected
48,157
Distribution
worldwide. Including United States, Virgin Islands and countries such as Argentina, Australia, Austria, Barbados, Belgium, Bermuda, Canada, Cayman Islands, Chile, Colombia, Cyprus, Czech Republic, Den
Date initiated
2005-06-24
Date posted
2005-06-30
Date terminated
2006-12-31
Location
Saint Paul, MN