ConMed Laparoscopic Instrument: Fenestrated (Duckbill) Disposable Grasper, 5mm x 33cm length, REF/Product Code 60-6045-0

FDA device recall Z-0101-2007 · posted 2006-10-27 · Terminated

Recall details

Recalling firm
Conmed Corporation
Reason for recall
The grasper jaws broke during laparoscopic procedures at the junction of the jaw and the tube.
Action taken
On 10/12/06, ConMed sent "URGENT DEVICE RECALL" packages (letter dated 10/11/06) to the domestic accounts via UPS Ground, and via UPS Express for the International accounts. Devices are to be removed from use and returned.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
GCJ
Quantity affected
970 units
Distribution
Worldwide
Date initiated
2006-10-12
Date posted
2006-10-27
Date terminated
2012-08-30
Location
Utica, NY