ConMed Goldline Hand Controlled With Needle Electrode, Rocker Switch, Safety Holster and 10 FT Cable, REF 130301A, Steri

FDA device recall Z-1562-2010 · posted 2010-05-10 · Terminated

Recall details

Recalling firm
ConMed Electrosurgery
Reason for recall
Electrosurgical pencil may remain activated or self activate after power switch is released.
Action taken
Customers were notified by letter on March 25, 2010, and were instructed to return all affected devices and to forward the recall information to anyone that may have received the devices to the end user. More information is available by contacting the firm directly at 800-552-0138.
Root cause
Device Design
Status
Terminated
Product code
GEI
Quantity affected
1400 units (US)
Distribution
Nationwide, including 2 VA facilities in GA and 1 military facility in UT. Foreign distribution to Australia, Belgium, Canada, China, Dominican Republic, El Salvador, Japan, Korea, Mexico, Puerto Rico
Date initiated
2010-03-29
Date posted
2010-05-10
Date terminated
2010-09-27
Location
Centennial, CO