Atlas FDA

CONFIRM anti-Progesterone Receptor (PR) (1E2) Rabbit Monoclonal Primary Antibody, Catalog Number 05278392001, model 790-

FDA device recall Z-3006-2018 · posted 2018-09-14 · Terminated

Recall details

Recalling firm
Ventana Medical Systems Inc
Reason for recall
Leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers. A false negative IHC result may occur.
Action taken
The firm disseminated their field safety notices by letter on 08/02/2018. The firm requested the destruction of the product.
Root cause
Under Investigation by firm
Status
Terminated
Product code
MXZ
Quantity affected
1655 units
Distribution
worldwide
Date initiated
2018-08-02
Date posted
2018-09-14
Date terminated
2021-11-22
Location
Oro Valley, AZ

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
CONFIRM ANTI-PROGESTORONE RECEPTOR (1E2) RABBIT MONOCLONAL PRIMARY ANTIBODY (K103818)Ventana Medical Sytems, Inc.2011-11-16