Computed Tomography X-Ray Systems (Brilliance 64, Model Number: 728231; Brilliance Big Bore, Model Number: 728243; Brill

FDA device recall Z-0845-2012 · posted 2012-01-24 · Terminated

Recall details

Recalling firm
Philips Medical Systems (Cleveland) Inc
Reason for recall
The Common Image Reconstruction System (CIRS) on the affected Brilliance Systems encounters an error and requires reboot of the system. The unit failed to initialize axial scan due to timeout that occurred waiting for "READY" state
Action taken
On 10/10/2011 the firm sent Urgent - Medical Device Correction letters to their customers. The letters identified the affected products, the problem, the hazard involved, the action to be taken by the customer, and actions planned by Philips. Customers were to reboot their systems if the two possible scenarios occurred. Philips Healthcare plans on installing the update through a Field Change Order free of charge. For further information or support concerning the issue, contact a local Philips representative or local Philips Healthcare office. In North America and Canada, contact Customer Care Solutions Center at 1-800-722-9377, option 5.
Root cause
Software design
Status
Terminated
Product code
JAK
Quantity affected
548 units
Distribution
Worldwide Distribution -- USA, including the states of: AL, AZ, CA, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NH, NM, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, VT,
Date initiated
2011-10-07
Date posted
2012-01-24
Date terminated
2013-06-18
Location
Cleveland, OH

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
BRILLIANCE VOLUME (K060937)Philips Medical Systems (Cleveland), Inc.2006-06-05
ACQSIM-MULTISLICE-CT, MODEL 453567042271 (K033357)Philips Medical Systems (Cleveland), Inc.2004-01-09
PHILIPS PLUS COMPUTED TOMOGRAPHY SYSTEM (K033326)Philips Medical Systems, Inc.2003-10-29