Computed Tomography X-Ray System (Brilliance CT Big Bore Oncology & Brilliance CT Big Bore Radiology) Distributed by Phi

FDA device recall Z-0801-2014 · posted 2014-01-21 · Terminated

Recall details

Recalling firm
Philips Medical Systems (Cleveland) Inc
Reason for recall
Machine Name in TumorLOC Basic Mode is obscured. When TumorLOC is in Basic Mode and Set Vertical Laser equals Yes, the Machine Name label and field are covered by the controls above it. The Machine Name cannot be viewed or changed.
Action taken
On 12/4/2013 the firm sent Urgent Medical Device Corrections letters to their customers.
Root cause
Software design
Status
Terminated
Product code
JAK
Quantity affected
32 units
Distribution
Worldwide Distribution-USA (nationwide) including the states of AK, MD, NH, NV, OH, and OR, and the countries of Australia, China, France, India, Indonesia, Italy, Korea, Mexico, Poland, Romania, Russ
Date initiated
2013-12-04
Date posted
2014-01-21
Date terminated
2015-07-01
Location
Cleveland, OH