Colleague CX Single Channel Volumetric Infusion Pumps, product code 2M8161and 2M8161R; Baxter Healthcare Corporation, De

FDA device recall Z-0344-05 · posted 2004-12-22 · Terminated

Recall details

Recalling firm
Baxter Healthcare Corp.
Reason for recall
A software anomaly causes a failure alarm code 12:303 which audibly alarms and stops the function of all channels in use. This causes an interruption in therapy, which may result in a risk to the patient.
Action taken
Baxter sent Urgent Device Correction letters dated 11/4/04 to the direct accounts to the attention of the Director of Biomedical Engineering on the same date. The letters informed the accounts of the 12:303 failure and possible interruption of patient therapy when it occurs. They were informed that new software was available to address the failure alarm code with the infusion pumps, and that the software also included updates to the predefined label library and the addition of a text message 'Close regulating roller clamp' to the manual tube release pop-up message. The software is being installed on all single and triple channel Colleague infusion pumps as they are processed through Baxter''s depot repair centers for other repair actions or routine maintenance. An Operator''s Manual Addendum was included with the letter to reflect the changes to the label library and the manual tube release pop-up message. The accounts were requested to complete the attached reply form and return it via fax to Baxter. The completed form acknowledges receipt of the recall letter, verifies the model numbers and serial numbers of the pumps at the facility and indicates the number of revised software installation kits needed by self-service customers. Any questions were directed to Baxter at 1-800-843-7867.
Root cause
Other
Status
Terminated
Product code
FRN
Distribution
Nationwide and internationally to Australia, the Bahamas, Belgium, Brazil, Canada, Colombia, El Salvador, Guatemala, Honduras, Israel, Lebanon, Mexico, New Zealand, Panama, the Philippines, Saudi Arab
Date initiated
2004-11-04
Date posted
2004-12-22
Date terminated
2011-02-02
Location
Round Lake, IL

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
COLLEAGUE (3CX, 3, CX) VOLUMETRIC INFUSION PUMP (K003616)Baxter Healthcare Corp2000-12-12
COLLEAGUE CX VOLUMETRIC INFUSION PUMP 2M8161 AND VOLUMETERIC INFUSION PUMP 2M8151 (K002211)Baxter Healthcare Corp2000-07-28