Atlas FDA

CollaPlug Absorbable Collagen Wound Dressing for Dental Surgery The Absorbable Collagen Wound Dressings for Dental Surge

FDA device recall Z-1293-2013 · posted 2013-05-10 · Terminated

Recall details

Recalling firm
Integra LifeSciences Corporation
Reason for recall
Due to the process deviation, product lots in question may have been released with higher levels of pyrogens than permitted by the specifications for the products.
Action taken
The firm, Integra, sent an "URGENT: MEDICAL DEVICE RECALL" letter dated April 9, 2013, to all US, Canada, and non-EU Integra consignees/customers. The letter described the product, problem and actions to be taken. The customers were instructed to verify if they have any lots of the listed product, if so, STOP using that product; remove the product from service; and promptly complete and return the PRODUCT RECALL RETURN ACKNOWLEDGMENT FORM via Email: FCA@integralife.com or FAX to : 1-609-275-9445. Customer Service department will contact you to return and replace the product. "However, if you have already implanted or used collagen sponge products affected by this recall, we recommend you monitor the patient for a fever in the immediate postoperative period according to the standard hospital or clinician protocol." For assistance or any other questions that you may have, please contact Customer Service at 1-855-532-1723.
Root cause
Process control
Status
Terminated
Product code
LPG
Quantity affected
155,850 boxes
Distribution
Worldwide distribution: USA (nationwide) including Puerto Rico and countries of: Arab Emirates, Argentina, Belgium, Brasil, Canada, Chile, China, Costa Rica, Denmark, Egypt, Spain, France, Great Brita
Date initiated
2013-03-28
Date posted
2013-05-10
Date terminated
2015-06-03
Location
Plainsboro, NJ