Codman ICP Skull Bolt Kit, Product Code: 82-6638. Single Use, disposable drill kit for gaining access to the cranium to

FDA device recall Z-1279-2012 · posted 2012-03-22 · Terminated

Recall details

Recalling firm
Codman & Shurtleff, Inc.
Reason for recall
Codman ICP Skull Bolt wrench is out of specification and cannot be used to turn the screw on the drill
Action taken
Codman issued "Urgent - Voluntary Product Recall Notification" letters on 2/9/12 via Fedx in the US and requested accounts examine inventory and return product. An Acknowledgement Form was requested to be returned. Outside the US affiliates were notified to recall to their accounts. Questions were directed to Customer Service (800) 225-0460.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
GWM
Quantity affected
4056 kits
Distribution
Worldwide Distribution - USA, and the countries of Canada, Austria, Brazil, Belgium, Columbia, China, Denmark, Estonia, Spain, Finland, France, Great Britain, Germany, India, Italy, Japan, Lithuania,
Date initiated
2012-02-09
Date posted
2012-03-22
Date terminated
2015-09-11
Location
Raynham, MA