CODMAN HAKIM BACTISEAL Shunt System Clear with Barium Stripe Peritoneal Catheter 120cm, with drainage Holes and Marking.
FDA device recall Z-1354-2021 · posted 2021-04-05 · Terminated
Recall details
- Recalling firm
- Integra LifeSciences Corp.
- Reason for recall
- Incorrect labeling, Catheter did not have the required holes and markings
- Action taken
- Integra LifeSciences issued Voluntary Urgent Medical Device Recall letter via courier service, facsimile or email on March 17th for delivery on March 19th, 2021. Letter states reason for recall, health risk and action to take: If you do have the affected product in Table 1, remove the product from service. b. Complete the attached acknowledgement form. If you do have affected product, check the box I do have affected product. Record the total quantity of the affected product you have. c. If you do not have affected product, check the box, I do not have affected product. d. Return the completed acknowledgement form to FCA2@integralife.com or FAX to 1-609-750-4220 e. Keep a copy of the form for your records. f. When your form is received, and it is noted that you have the affected product, Customer Service will contact you and provide a Return Material Authorization (RMA) number and directions to return the product. A replacement order will also be placed for the quantity noted on the form. Receipt of the acknowledgement form ensures that Integra has achieved a level of effectiveness in communicating this information effectively. We recommend you also maintain a copy of this notification and signed copy of the acknowledgement forms for your records. Regulatory agencies such as the FDA perform audits of field actions of this nature to verify that our customers have been notified and understand the nature of the field action. Should you have any questions regarding these instructions, please contact Customer Service: USA: 800-654-2873 Monday- Friday 8:00AM EST 6:00PM EST custsvcnj@integralife.com
- Root cause
- Labeling Change Control
- Status
- Terminated
- Product code
- HCA
- Quantity affected
- 49 units
- Distribution
- CA CO FL KY MD NC NC NY NY TX WI
- Date initiated
- 2021-03-17
- Date posted
- 2021-04-05
- Date terminated
- 2021-12-02
- Location
- Princeton, NJ